|| TYPICAL ROOM LAYOUT
RadPRO® OTX
Overhead Radiography System
     
     
     
     
     

 

|| EQUIPMENT SCHEDULE
|| GENERAL NOTES
 
Equipment Identifier
 
BTU / HR
WEIGHT (lbs.)
   
Description
DIMENSIONS (in)
W x D x H
 
 
334
RADPRO OTX
VPC SERIES 50kW HF GENERATOR
CONTROL CONSOLE
16.60
12.30
3.70
4
 
1337
RADPRO OTX
VPC SERIES 50kW HF GENERATOR
POWER CABINET
18.00
16.00
48.70
340
 
400
RADPRO OTX
VPC SERIES ELEVATOR 2
ELEVATING 4-WAY FLOAT TABLE
86.62
31.89
33.47
772
 
1200
RADPRO OTX
VST SERIES 3D TOP
CEILING MOUNTED TUBE SUPPORT
168.00
118.50
112.79
1200
 
35
RADPRO OTX
VPC SERIES BS-2000V
NON-TILTING WALLSTAND
25.60
16.54
87.40
330
         
         

All specifications and space requirements are shown generically. Contact virtual imaging for specific room requirements pertaining to optional equipment components and additional room preparation specifications

Absolute minimum room dimensions:
16’-0” w x 14’-0” d

Minimum recommended ceiling height: 9’-0”

Suggested ambient room requirements:
66 to 75 degrees F at 30% to 60% r.m. (Non-condensing).

Power option #1:
Three phase power

Main power to generator:
480 vac, 3-phase, 70kva

Auxiliary power:
208/120 vac, 30 amp load center
(No other options are available for this generator)