The RadPRO® OT2 Overhead Radiography system is perfect for busy hospitals and outpatient centers.. This system is equipped standard with a high frequency generator (100kHz), four-way float table (500 lbs capacity) and heavy duty overhead tube crane that all®ows for easy, efficient exams needed by today's high volume hospital imaging departments.
Heavy Duty Overhead Tube Crane w/ Motorized Vertical Movement
•
Low Absorption Tabletop
•
Fully Upgradeable with Canon CXDI DR Imaging Systems
||SPECIFICATIONS
Generator
VIEW
High Freguency Generator
100kHz
Power Output
55 kW, 65 kW, 80 kW
High Speed Starter
Yes (Standard)
Automatic Exposure Control
Yes (Standard)
Automatic Programming
Yes, with easy programming on standard touch screen console
Overhead Tube Crane
VIEW
Travel - Longitudinal
139.3" in (354 cm)
Travel - Tranverse
87.4" in (222 cm)
Travel - Vertical
59" in (150 cm)
Focus - Ceiling Distance
34.4" in (87.4 cm)
Focus - Floor Distance
17.7" to 74.8" in (45 to 190 cm)
Tube Rotation - Vertical Axis
+154°, -182°, Detents Every 90°
Tube Support - Travel
63” in (160 cm)
Tube Support - Rotation
+90°, -90°
X-Ray Tube
300 kHu, 400 kHu, 600 kHu
Manual Collimator (VST-MC)
Rectangular Shaped Collimation
Scales for SID’s 36”, 40”, 48”, 72” in
Rotation by 45° degrees with a stop in 0° position by swiveling flange
Light localizer lamp w/ time switch 30 sec.
Elevating Table ( 4-Way Float )
VIEW
Maximum Patient Weight
500 lbs (272 kg)
Tabletop - Width
31.5" in (80 cm)
Tabletop - Length
94.5" in (240 cm)
Table Height
Elevating from 23.2” to 35” in, ± 0.6” in
(59 cm to 89 cm ± 1.5 cm)
Tabletop Travel - Longitudinal
± 18.9" in (± 48 cm)
Tabletop Travel - Tranverse
± 5.5" in (± 14 cm)
Tabletop - Film Distance
2.08" in, ±.08" in (53 mm, ± 2 mm)
Radiation Absorption
</= 0.6 mm AI
Wall / Chest Stand
VIEW
Travel - Vertical
53" in (134 cm)
Central Beam Height
15" to 74.8" in ( 38 cm to 190 cm)
Single Handed Cassette/Detector Tray
Accepts 5" x 7" in / 14" x17" in Cassettes/Dectectors
Digital Detector (Optional)
VIEW
Canon CXDI-50 Series (14" x17")
Models: CXDI-50G, CXDI-50C
Canon CXDI-40 Series (17" x 17")
Models: CXDI-40EG, CXDI-40EC
Snap-On Grid (Optional)
Compatible w/ CXDI-50G, CXDI-50C Models
DR50 Digital Receptor (Optional)
VIEW
Designed with the Canon CXDI-50 Series in mind, the RadPRO® DR50 Digital Receptor can manually be adjusted to portrait or landscape exposures in one smooth motion and without having to remove the digital flat panel.
||Canon CXDI Technology
||Canon Software Solutions
Models: CXDI-50G, CXDI-50C
Models:CXDI-40EG, CXDI-40EC
DR Image Viewer
Program Image mode
Specfications are subject to change without notice.
Certain configurations may not be available. Contact your local authorized RadPRO® or Canon representative for information on specific configurations.
Names of companies or products appearing on this website or in supporting documentation provided by this website are trademarks and/or registered
trademarks of their respective owners. Canon and CXDI are registered trademarks of Canon, Inc. in the United States and may also be registered
trademarks or trademarks in other countries.
RadPRO® is a division and registered trademark of Virtual Imaging, Inc. All rights reserved.