Digital Radiography Premium Flat Panel Detector
 

Introducing the CXDI-40EC, a premium DR system that offers fast, filmless X-Ray capture for diverse radiographic applications. Featuring Canon's advanced flat panel detector technology with Cesium Iodide (CsI) scintillator for superb dose efficiency, the high-sensitivity detector delivers clear, high-resolution images with lower X-ray dosage than conventional systems. In addition to offering easier patient positioning, seamless network integration, and
significantly reduced wait time for both patients and
staff, the CXDI-40EC can also be flexibly installed in
various configurations to fit your exact requirements.

 
|| FEATURES
Large Imaging Area: 17 x 17 in (43 x 43 cm)
High Quality Diagnostic Images: Amorphous Silicon (a-Si)
Immediate Results: 3-5 Second Image Acquisition After Exposure
Diverse Applications: Existing Table, Wall Stand or Universal Stand
DICOM Compliant
Seamleas Network Integration: Ethernet 10/100 Base T Connectivity
   
 
 






|| SPECIFICATIONS
CXDI-40EC
VIEW
Purpose
General Radiography
Method
Flat Panel Detector:
Scintillator & Amorphous Silicon (a-Si)
Sensor
LANMIT 8 ( Large Area New-MIS sensor and TFT)
Scintillator
CsI (Cesium Iodide): TI
Pixel Pitch
160 x 160 microns
Pixels
2,688 x 2,688 pixels (7.2 million pixels)
Image Size
Automatic sizing up 17" x 17" in (43 x43 cm)
A/D
14-bit
Grayscale
4,096 grayscale (12-bit)
Preview Image Acqisition Time*
Approx. 3-5 seconds after X-Ray exposure
Interface
DICOM 3.0, Ethernet 10/100 Base T
DICOM
DICOM 3.0 Compatible
Print Management Service Class (SCU)
Storage Service Class (SCU), and others.
Voltage
100V, 120V, 230/240V (50/60Hz)
Power Consumption
Sensor Unit: 200VA maximum
Operating Environment
Sensor Unit: 41°- 95° F (5°-35° C)
30-75% RH (non-condensing)
Certification
FDA 510(k), IEC60601-1-2:2001 Class A, UL 60601-1,
EN60601-1, CE0197
Dimensions
Sensor Unit: 21.7 x 21.7 x 2.7 in (500 x 550 x 67.5)
Weight
Sensor Unit: 44lbs (20kg), without grid
 
* Actual times may differ due to various factors.
User Options
VIEW
Grid
Please contact your local authorized Canon rep.
Software Options
Please contact your local authorized Canon rep.
System Configuration (example #1)
VIEW
For Exact system configuration details, please contact your local authorized RadPRO® or Canon representative.
 
System Configuration (example #2)
VIEW
For Exact system configuration details, please contact your local authorized RadPRO® or Canon representative.
 
System Configuration (example #3)
VIEW
For Exact system configuration details, please contact your local authorized RadPRO® or Canon representative.
 
 
 
 

|| Canon In Action

 
 
 
 
 
 
 
 
 
 

Specfications are subject to change without notice.

Certain configurations may not be available. Contact your local authorized RadPRO® or Canon representative for information on specific configurations.

Names of companies or products appearing on this website or in supporting documentation provided by this website are trademarks and/or registered
trademarks of their respective owners. Canon and CXDI are registered trademarks of Canon, Inc. in the United States and may also be registered
trademarks or trademarks in other countries.

RadPRO® is a division and registered trademark of Virtual Imaging, Inc. All rights reserved.
 
 
 
© 2004-2008 Virtual Imaging, Inc. All Rights Reserved.